        AusCERT Security Bulletin Summary
        ---------------------------------

Product:           Medtronic 2090 Carelink Programmer
Publisher:         ICS-CERT
Operating System:  Network Appliance
Impact/Access:     Access Privileged Data -- Console/Physical
                   Read-only Data Access  -- Existing Account
Resolution:        Patch/Upgrade
CVE Names:         CVE-2018-5448 CVE-2018-5446 

Original Bulletin: 
   https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01

Revision History:  January  31 2020: Updated bulletin from ICS-CERT re Vendor patches
                   February 28 2018: Initial Release

- --------------------------BEGIN INCLUDED TEXT--------------------

ICS Advisory (ICSMA-18-058-01)

Medtronic 2090 Carelink Programmer Vulnerabilities (Update C)

Original release date: January 30, 2020

Legal Notice

All information products included in https://us-cert.gov/ics are provided"as
is" for informational purposes only. The Department of Homeland Security (DHS)
does not provide any warranties of any kind regarding any information contained
within. DHS does not endorse any commercial product or service, referenced in
this product or otherwise. Further dissemination of this product is governed by
the Traffic Light Protocol (TLP) marking in the header. For more information
about TLP, see https://www.us-cert.gov/tlp/ .



1. EXECUTIVE SUMMARY

  o CVSS v3 7.1
  o Vendor: Medtronic
  o Equipment: 2090 CareLink Programmer, 29901 Encore Programmer
  o Vulnerabilities: Storing Passwords in a Recoverable Format, Relative Path
    Traversal, Improper Restriction of Communication Channel to Intended
    Endpoints

2. UPDATE INFORMATION

This updated advisory is a follow-up to the updated advisory titled
ICSMA-18-058-01 Medtronic 2090 CareLink Programmer Vulnerabilities (Update B)
that was published October 11, 2018, ICS webpage on us-cert.gov.

3. RISK EVALUATION

Successful exploitation of these vulnerabilities may allow an attacker with
physical access to a 2090 Programmer to obtain per-product credentials to the
software deployment network. These credentials grant access to the software
deployment network, but access is limited to read-only versions of device
software applications.

Additionally, successful exploitation of these vulnerabilities may allow an
attacker with local network access to influence communications between the
Programmer and the software deployment network.

Additional product security review revealed potential for an attacker to
remotely exploit a subset of these vulnerabilities. The SessionSync and
RemoteView features associated with the affected products are not affected by
the vulnerabilities reported in this advisory.

4. TECHNICAL DETAILS

4.1 AFFECTED PRODUCTS

The following versions of the Medtronic CareLink 2090 Programmer, a device used
by trained personnel at hospitals and clinics to program and manage Medtronic
cardiac devices, are affected:

  o 2090 CareLink Programmer, all versions
  o 29901 Encore Programmer, all versions

4.2 VULNERABILITY OVERVIEW

4.2.1 STORING PASSWORDS IN A RECOVERABLE FORMAT CWE-257

The affected product uses a per-product username and password that is stored in
a recoverable format.

CVE-2018-5446 has been assigned to this vulnerability. A CVSS v3 base score of
4.9 has been calculated; the CVSS vector string is ( AV:P/AC:H/PR:N/UI:N/S:C/
C:H/I:N/A:N ).

4.2.2 RELATIVE PATH TRAVERSAL CWE-23

The affected product's software deployment network contains a directory
traversal vulnerability that could allow an attacker to read files on the
system.

CVE-2018-5448 has been assigned to this vulnerability. A CVSS v3 base score of
4.8 has been calculated; the CVSS vector string is ( AV:A/AC:H/PR:L/UI:N/S:U/
C:H/I:N/A:N ).

4.2.3 IMPROPER RESTRICTION OF COMMUNICATION CHANNEL TO INTENDED ENDPOINTS
CWE-923

The affected product uses a virtual private network connection to securely
download updates. The product does not verify it is still connected to this
virtual private network before downloading updates. The affected products
initially establish an encapsulated IP-based VPN connection to a
Medtronic-hosted update network. Once the VPN is established, it makes a
request to a HTTP (non-TLS) server across the VPN for updates, which responds
and provides any available updates. The programmer-side (client) service
responsible for this HTTP request does not check to ensure it is still
connected to the VPN before making the HTTP request. Thus, an attacker could
cause the VPN connection to terminate (through various methods and attack
points) and intercept the HTTP request, responding with malicious updates via a
man-in-the-middle attack. The affected products do not verify the origin or
integrity of these updates, as it insufficiently relied on the security of the
VPN. An attacker with remote network access to the programmer could influence
these communications.

CVE-2018-10596 has been assigned to this vulnerability. A CVSS v3 base score of
7.1 has been calculated; the CVSS vector string is. ( AV:N/AC:H/PR:L/UI:R/S:U/
C:H/I:H/A:H ).

4.3 BACKGROUND

  o CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  o COUNTRIES/AREAS DEPLOYED: Worldwide
  o COMPANY HEADQUARTERS LOCATION: Ireland

4.4 RESEARCHER

Researchers Billy Rios and Jonathan Butts of Whitescope LLC identified these
vulnerabilities and reported them to CISA.

5. MITIGATIONS

Medtronic has assessed the vulnerabilities and determined that no new potential
safety risks were identified. In order to enhance system security, Medtronic
has added periodic integrity checks for files associated with the software
deployment network. Additionally, Medtronic has developed server-side security
changes that further enhance security. Medtronic reports that they will not be
issuing a product update; however, Medtronic has identified compensating
controls within this advisory to reduce the risk of exploitation and reiterates
the following from the CareLink 2090 Programmer Reference Manual:

  o Maintain good physical controls over the programmer. Having a secure
    physical environment prevents access to the internals of the programmer.
  o Only connect the programmer to managed, secure networks.
  o Update the software on the programmer when Medtronic updates are available.
  o Alternatively, disconnect the programmer from the network. Network
    connectivity is not required for normal programmer operation.
  o Offline updates are available, contact your Medtronic representative for
    more information.

- --------- Begin Update C Part 1 of 1 --------

Medtronic has deployed mitigating patches to address the reported
vulnerabilities. Medtronic has also stated that they have increased security
controls associated with these vulnerabilities. As a result of the available
mitigating patches, Medtronic has re-enabled the network-based software update
mechanism.

Medtronic has stated that the patch for affected products can be obtained by
contacting Medtronic Technical Services at 800-638-1991.

- --------- End Update C Part 1 of 1 --------

After additional review and risk evaluation of the affected products, Medtronic
has disabled the network-based software update mechanism, including both the
VPN and the HTTP subservices, as an immediate security mitigation. Users should
not attempt to update the affected products over the network as this update
mechanism is vulnerable to the attack described in section 4.2.3. Medtronic
will continue to implement and deploy increased security protections and
mitigations to address the vulnerabilities in this advisory.

Users should still obtain and apply updates via controlled USB dongles and
should contact their Medtronic representative for more information.

Medtronic recommends that affected products continue to be used for their
intended purpose in the previously described manner.
Medtronic has released a security bulletin for the 2090 CareLink Programmer .

CISA reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices
on the ICS webpage on us-cert.gov . Several recommended practices are available
for reading and download, including Improving Industrial Control Systems
Cybersecurity with Defense-in-Depth Strategies .

Additional mitigation guidance and recommended practices are publicly available
on the ICS webpage on us-cert.gov in the Technical Information Paper,
ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation
Strategies .

Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to CISA for tracking
and correlation against other incidents.

These vulnerabilities are not exploitable remotely.

For any questions related to this report, please contact the CISA at:

Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870

CISA continuously strives to improve its products and services. You can help by
choosing one of the links below to provide feedback about this product.

- --------------------------END INCLUDED TEXT--------------------

You have received this e-mail bulletin as a result of your organisation's
registration with AusCERT. The mailing list you are subscribed to is
maintained within your organisation, so if you do not wish to continue
receiving these bulletins you should contact your local IT manager. If
you do not know who that is, please send an email to auscert@auscert.org.au
and we will forward your request to the appropriate person.

NOTE: Third Party Rights
This security bulletin is provided as a service to AusCERT's members.  As
AusCERT did not write the document quoted above, AusCERT has had no control
over its content. The decision to follow or act on information or advice
contained in this security bulletin is the responsibility of each user or
organisation, and should be considered in accordance with your organisation's
site policies and procedures. AusCERT takes no responsibility for consequences
which may arise from following or acting on information or advice contained in
this security bulletin.

NOTE: This is only the original release of the security bulletin.  It may
not be updated when updates to the original are made.  If downloading at
a later date, it is recommended that the bulletin is retrieved directly
from the author's website to ensure that the information is still current.

Contact information for the authors of the original document is included
in the Security Bulletin above.  If you have any questions or need further
information, please contact them directly.

Previous advisories and external security bulletins can be retrieved from:

        https://www.auscert.org.au/bulletins/

===========================================================================
Australian Computer Emergency Response Team
The University of Queensland
Brisbane
Qld 4072

Internet Email: auscert@auscert.org.au
Facsimile:      (07) 3365 7031
Telephone:      (07) 3365 4417 (International: +61 7 3365 4417)
                AusCERT personnel answer during Queensland business hours
                which are GMT+10:00 (AEST).
                On call after hours for member emergencies only.
===========================================================================
-----BEGIN PGP SIGNATURE-----
Comment: http://www.auscert.org.au/render.html?it=1967
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=Tsh3
-----END PGP SIGNATURE-----