-----BEGIN PGP SIGNED MESSAGE-----
Hash: SHA256

===========================================================================
             AUSCERT External Security Bulletin Redistribution

                               ESB-2022.5282
        Advisory (icsma-20-296-02) B. Braun SpaceCom, Battery Pack
           SP with Wi-Fi, and Data module compactplus (Update A)
                              21 October 2022

===========================================================================

        AusCERT Security Bulletin Summary
        ---------------------------------

Product:           B. Braun SpaceCom
                   attery Pack SP with Wi-Fi
                   Data module compactplus
Publisher:         ICS-CERT
Operating System:  Network Appliance
Resolution:        Patch/Upgrade
CVE Names:         CVE-2020-25168 CVE-2020-25166 CVE-2020-25164
                   CVE-2020-25162 CVE-2020-25160 CVE-2020-25158
                   CVE-2020-25156 CVE-2020-25154 CVE-2020-25152
                   CVE-2020-25150 CVE-2020-16238 

Original Bulletin: 
   https://us-cert.cisa.gov/ics/advisories/icsma-20-296-02

Comment: CVSS (Max):  7.6 CVE-2020-25158 (CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:L/A:L)
         CVSS Source: ICS-CERT
         Calculator:  https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:L/A:L

- --------------------------BEGIN INCLUDED TEXT--------------------

ICS Medical Advisory (ICSMA-20-296-02)

B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
(Update A)

Original release date: October 22, 2022 | Last revised: October 20, 2022

Legal Notice

All information products included in https://us-cert.cisa.gov/ics are provided
"as is" for informational purposes only. The Department of Homeland Security
(DHS) does not provide any warranties of any kind regarding any information
contained within. DHS does not endorse any commercial product or service,
referenced in this product or otherwise. Further dissemination of this product
is governed by the Traffic Light Protocol (TLP) marking in the header. For more
information about TLP, see https://us-cert.cisa.gov/tlp/ .



1. EXECUTIVE SUMMARY

  o CVSS v3 7.6
  o ATTENTION: Exploitable remotely/low and high skill level to exploit
  o Vendor: B. Braun Melsungen AG
  o Equipment: SpaceCom, Battery Pack SP with Wi-Fi, and Data module
    compactplus
  o Vulnerabilities : Cross-site Scripting, Open Redirect, XPath Injection,
    Session Fixation, Use of a One-way Hash without a Salt, Relative Path
    Traversal, Improper Verification of Cryptographic Signature, Improper
    Privilege Management, Use of Hard-coded Credentials, Active Debug Code,
    Improper Access Control

2. UPDATE

This updated advisory is a follow-up to the original advisory titled
"ICSMA-20-296-02 B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module
compactplus" that was published October 22, 2020, on the ICS webpage on
cisa.gov/ICS.

3. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to
compromise the security of the Space or compactplus communication devices,
allowing an attacker to escalate privileges, view sensitive information, upload
arbitrary files, and perform remote code execution.

4. TECHNICAL DETAILS

4.1 AFFECTED PRODUCTS

The following versions of B. Braun products are affected:

  o SpaceCom, software Versions U61 and earlier (United States), L81 and
    earlier (outside the United States)
  o Battery pack with Wi-Fi, software Versions U61 and earlier (United States),
    L81 and earlier (outside the United States)
  o Data module compactplus, software Versions A10 and A11 (not distributed in
    the United States)

4.2 VULNERABILITY OVERVIEW

4.2.1 IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION ('CROSS-SITE
SCRIPTING') CWE-79

A reflected cross-site scripting (XSS) vulnerability in the B. Braun Melsungen
AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus
Versions A10 and A11 allows remote attackers to inject arbitrary web script or
HTML into various locations.

CVE-2020-25158 has been assigned to this vulnerability. A CVSS v3 base score of
7.6 has been assigned; the CVSS vector string is ( AV:N/AC:L/PR:N/UI:R/S:U/C:H/
I:L/A:L ).

4.2.2 URL REDIRECTION TO UNTRUSTED SITE ('OPEN REDIRECT') CWE-601

An open redirect vulnerability in the administrative interface of the B. Braun
Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module
compactplus Versions A10 and A11 allows attackers to redirect users to
malicious websites.

CVE-2020-25154 has been assigned to this vulnerability. A CVSS v3 base score of
5.4 has been assigned; the CVSS vector string is ( AV:N/AC:L/PR:N/UI:R/S:U/C:L/
I:L/A:N ).

4.2.3 IMPROPER NEUTRALIZATION OF DATA WITHIN XPATH EXPRESSIONS ('XPATH
INJECTION') CWE-643

A XPath injection vulnerability in the B. Braun Melsungen AG SpaceCom Version
L81/U61 and earlier, and the Data module compactplus Versions A10 and A11
allows unauthenticated remote attackers to access sensitive information and
escalate privileges.

CVE-2020-25162 has been assigned to this vulnerability. A CVSS v3 base score of
7.5 has been assigned; the CVSS vector string is ( AV:N/AC:L/PR:N/UI:N/S:U/C:H/
I:N/A:N ).

4.2.4 SESSION FIXATION CWE-384

A session fixation vulnerability in the B. Braun Melsungen AG SpaceCom
administrative interface Version L81/U61 and earlier, and the Data module
compactplus Versions A10 and A11 allows remote attackers to hijack web sessions
and escalate privileges.

CVE-2020-25152 has been assigned to this vulnerability. A CVSS v3 base score of
6.5 has been assigned; the CVSS vector string is ( AV:N/AC:L/PR:N/UI:R/S:U/C:H/
I:N/A:N ).

4.2.5 USE OF A ONE-WAY HASH WITHOUT A SALT CWE-759

A vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and
earlier, and the Data module compactplus Versions A10 and A11 allows attackers
to recover user credentials of the administrative interface.

CVE-2020-25164 has been assigned to this vulnerability. A CVSS v3 base score of
5.1 has been assigned; the CVSS vector string is ( AV:L/AC:H/PR:N/UI:N/S:U/C:H/
I:N/A:N ).

4.2.6 RELATIVE PATH TRAVERSAL CWE-23

A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version
L81/U61 and earlier, and the Data module compactplus Versions A10 and A11
allows attackers with service user privileges to upload arbitrary files. By
uploading a specially crafted tar file an attacker can execute arbitrary
commands.

CVE-2020-25150 has been assigned to this vulnerability. A CVSS v3 base score of
7.6 has been assigned; the CVSS vector string is ( AV:N/AC:L/PR:L/UI:N/S:U/C:L/
I:H/A:L ).

4.2.7 IMPROPER VERIFICATION OF CRYPTOGRAPHIC SIGNATURE CWE-347

An improper verification of the cryptographic signature of firmware updates of
the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data
module compactplus Versions A10 and A11 allows attackers to generate valid
firmware updates with arbitrary content that can be used to tamper with
devices.

CVE-2020-25166 has been assigned to this vulnerability. A CVSS v3 base score of
6.8 has been assigned; the CVSS vector string is ( AV:P/AC:H/PR:N/UI:N/S:C/C:N/
I:H/A:H ).

4.2.8 IMPROPER PRIVILEGE MANAGEMENT CWE-269

A vulnerability in the configuration import mechanism of the B. Braun Melsungen
AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus
Versions A10 and A11 allows attackers with command line access to the
underlying Linux system to escalate privileges to the root user.

CVE-2020-16238 has been assigned to this vulnerability. A CVSS v3 base score of
6.7 has been assigned; the CVSS vector string is ( AV:L/AC:L/PR:H/UI:N/S:U/C:H/
I:H/A:H ).

4.2.9 USE OF HARD-CODED CREDENTIALS CWE-798

Hard-coded credentials in the B. Braun Melsungen AG SpaceCom Version L81/U61
and earlier, and the Data module compactplus Versions A10 and A11 enables
attackers with command line access to access the device's Wi-Fi module.

CVE-2020-25168 has been assigned to this vulnerability. A CVSS v3 base score of
3.3 has been assigned; the CVSS vector string is ( AV:L/AC:L/PR:L/UI:N/S:U/C:L/
I:N/A:N ).

4.2.10 ACTIVE DEBUG CODE CWE-489

Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the
Data module compactplus Versions A10 and A11 and earlier enables attackers in
possession of cryptographic material to access the device as root.

CVE-2020-25156 has been assigned to this vulnerability. A CVSS v3 base score of
7.2 has been assigned; the CVSS vector string is ( AV:N/AC:L/PR:H/UI:N/S:U/C:H/
I:H/A:H ).

4.2.11 IMPROPER ACCESS CONTROL CWE-284

Improper access controls in the B. Braun Melsungen AG SpaceCom Version L81/U61
and earlier, and the Data module compactplus Versions A10 and A11 enables
attackers to extract and tamper with the devices network configuration.

CVE-2020-25160 has been assigned to this vulnerability. A CVSS v3 base score of
6.8 has been assigned; the CVSS vector string is ( AV:L/AC:L/PR:N/UI:N/S:C/C:L/
I:L/A:L ).

4.3 BACKGROUND

  o CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  o COUNTRIES/AREAS DEPLOYED: Worldwide
  o COMPANY HEADQUARTERS LOCATION: Germany

4.4 RESEARCHER

Julian Suleder, Nils Emmerich, and Birk Kauer of ERNW Research GmbH; Dr. Oliver
Matula of ERNW Enno Rey Netzwerke GmbH, reported these vulnerabilities to the
Federal Office for Information Security (BSI), Germany, in the context of the
BSI project ManiMed (Manipulation of medical devices).

5. MITIGATIONS

- --------- Begin Update A part 1 of 1 ---------

B. Braun has released software updates to mitigate the reported
vulnerabilities:

Within the United States and Canada:

Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower).
Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher).
SpaceStation with SpaceCom 2, software Version 012U000093.

For details on acquiring this software, see the B. Braun Advisory .

Users in the United States and Canada who need additional support can contact
B. Braun Technical Support by calling 800-627-PUMP or by emailing
AISTechSupport@bbraunusa.com .

Note: Facilities in Canada using "U" versions of software should follow the
U.S. vulnerability disclosure. Facilities in Canada using non "U" versions
(e.g. L) should follow the vulnerability disclosure for outside the U.S.

Outside the United States and Canada:

  o Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853).
  o Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher).
  o SpaceStation with SpaceCom 2, software version 011L000093.

For more information, see the B. Braun's Vulnerability Advisory .

- --------- End Update A part 1 of 1 ---------

In addition, B. Braun recommends users of the affected products consider the
following best practices:

  o All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should
    review their IT infrastructure to ensure a network zone concept has been
    implemented whereby critical systems, such as infusion pumps, are housed in
    separate (e.g., by firewalls or VLAN) environments that are not accessible
    directly from the Internet or by unauthorized users.
  o Wireless networks should be implemented using industry standard encryption
    and should be equipped with intrusion detection systems (IDS) and/or
    intrusion prevention systems (IPS).

CISA recommends users take defensive measures to minimize the risk of
exploitation of this vulnerability. Specifically, users should:

  o Minimize network exposure for all control system devices and/or systems,
    and ensure that they are not accessible from the Internet .
  o Locate control system networks and remote devices behind firewalls, and
    isolate them from the business network.
  o When remote access is required, use secure methods, such as Virtual Private
    Networks (VPNs), recognizing that VPNs may have vulnerabilities and should
    be updated to the most current version available. Also recognize that VPN
    is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices
on the ICS webpage on us-cert.cisa.gov . Several recommended practices are
available for reading and download, including Improving Industrial Control
Systems Cybersecurity with Defense-in-Depth Strategies .

Additional mitigation guidance and recommended practices are publicly available
on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper,
ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation
Strategies .

Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to CISA for tracking
and correlation against other incidents.

No known public exploits specifically target these vulnerabilities. These
vulnerabilities are exploitable remotely. These vulnerabilities have a low
attack complexity.

For any questions related to this report, please contact the CISA at:

Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870

CISA continuously strives to improve its products and services. You can help by
choosing one of the links below to provide feedback about this product.

- --------------------------END INCLUDED TEXT--------------------

You have received this e-mail bulletin as a result of your organisation's
registration with AusCERT. The mailing list you are subscribed to is
maintained within your organisation, so if you do not wish to continue
receiving these bulletins you should contact your local IT manager. If
you do not know who that is, please send an email to auscert@auscert.org.au
and we will forward your request to the appropriate person.

NOTE: Third Party Rights
This security bulletin is provided as a service to AusCERT's members.  As
AusCERT did not write the document quoted above, AusCERT has had no control
over its content. The decision to follow or act on information or advice
contained in this security bulletin is the responsibility of each user or
organisation, and should be considered in accordance with your organisation's
site policies and procedures. AusCERT takes no responsibility for consequences
which may arise from following or acting on information or advice contained in
this security bulletin.

NOTE: This is only the original release of the security bulletin.  It may
not be updated when updates to the original are made.  If downloading at
a later date, it is recommended that the bulletin is retrieved directly
from the author's website to ensure that the information is still current.

Contact information for the authors of the original document is included
in the Security Bulletin above.  If you have any questions or need further
information, please contact them directly.

Previous advisories and external security bulletins can be retrieved from:

        https://www.auscert.org.au/bulletins/

===========================================================================
Australian Computer Emergency Response Team
The University of Queensland
Brisbane
Qld 4072

Internet Email: auscert@auscert.org.au
Facsimile:      (07) 3365 7031
Telephone:      (07) 3365 4417 (International: +61 7 3365 4417)
                AusCERT personnel answer during Queensland business hours
                which are GMT+10:00 (AEST).
                On call after hours for member emergencies only.
===========================================================================
-----BEGIN PGP SIGNATURE-----
Comment: https://auscert.org.au/gpg-key/
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=ZsMz
-----END PGP SIGNATURE-----